User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: HP 5315A SUB PROCEDURE 1 DATE: 23-Oct-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 3 NUMBER OF LINES: 56 CONFIGURATION: HP 3325B STANDARD: Stanford Research FS700 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F 1.002 ASK+ X 1.003 DISP Allow the UUT a 1 hour warm-up. 1.004 HEAD {* FREQUENCY ACCURACY TEST *} 1.005 DISP Set UUT front panel controls as follows: 1.005 DISP 1.005 DISP FUNCTION SELECT switch .............. FREQ C/GATE TIME 1.005 DISP FUNCTION SET (Blue shift key) ....... IN 1.005 DISP GATE TIME/DELAY control ............. Adjust 1.005 DISP CHANNELS A&B: 1.005 DISP TRIG LVL/SENS controls .............. MAX (CW) 1.005 DISP TRIGGER LEVEL/SENSITIVITY ........... SENS (IN) 1.006 DISP Set UUT front panel controls as follows: 1.006 DISP 1.006 DISP CHANNELS A&B: 1.006 DISP SLOPES .............................. POS (OUT) 1.006 DISP AC/DC ................................ DC (IN) 1.006 DISP ATTN x1/x20 .......................... x1 (OUT) 1.006 DISP FILTER NORM/100 kHz .................. FILTER NORM (OUT) 1.006 DISP SEP/COM A ............................ SEP (OUT) 1.007 DISP Adjust GATE TIME control for a 1.007 DISP UUT reading of 1.00 S ± .05 S. 1.008 DISP Set UUT front panel controls as follows: 1.008 DISP 1.008 DISP FUNCTION SELECT switch .............. FREQ A/PER A 1.008 DISP FUNCTION SET (Blue shift key) ....... OUT 1.009 DISP Connect the 3325A output to the UUT INPUT A 1.009 DISP through a 50 ohm feedthrough. 1.010 STD Stanford Research FS700 1.011 JMP 2.001 1.012 EVAL 2.001 IEEE [@3325] FU 1; AM 1 VR; FR 10 MH 2.002 MEMI Use keyboard to enter actual UUT reading in MHz. 2.003 ACC 10.000000MH 0.0000009U 2.004 MEME 2.005 MEMC MH 0.000046U #! Test Tol 4.6e-005, Sys Tol 9e-007, TUR 51.111 (>= 4.00). 3.001 END #! T.U.R.s less than 4.00: 0 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 0 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.